REMOTE CONTROL
141
Command
:SOURce[1|2]:VOLTage?
Function
Queries the output voltage setting.
Example
:SOURce2:VOLTage?
Queries the output voltage setting on CH2.
Command
VSET<X>?
Function
Returns the output voltage setting.
Description
X 1= CH1, 2= CH2
Example
VSET1?
Returns the CH1 output voltage setting.
Command
:SOURce[1|2]:RESistor < NRf >
Function
Set the resistor value.
Description
NRF
1-1000
Example
:SOURce2:RESistor 1000
Set the resistor value as 1000Ω for CH2.
Command
:SOURce[1|2]:RESistor?
Function
Queries the setting resistor value.
Example
:SOURce2:RESistor?
Queries the setting resistor value and returns the
readback value for CH2.
Command
:SOURce:CURRent:ALL?
Function
Queries the setting current value in all channnels.