REMOTE CONTROL
169
Example
:SEQUence2:TEMPlet:INVErt?
Returns to the templet of CH2 to check if the invert
function is ON or OFF.
Command
:SEQUence[1|2]:TEMPlet:MAXValue{<value>|MI
Nimum|MAXimum}
Function
Sets the maximum voltage or current value of the
active selected templet.
Description
When voltage is the object to edit, it is the
maximum voltage to be set. When current is the
object to edit, it is the maximum current to be set.
When templet type is Pulse, the command is to set
low level value.
Example
:SEQUence2:TEMPlet:MAXValue 5
Sets the CH2 maximum voltage of the active
selected templet as 5V.
Command
:SEQUence[1|2]:TEMPlet:MAXValue?
Function
Queries the maximum voltage or current value of
the active selected templet.
Example
:SEQUence2:TEMPlet:MAXValue?
Queries the CH2 maximum voltage or current
value of the active selected templet.
Command
:SEQUence[1|2]:TEMPlet:MINValue{<value>|MI
Nimum|MAXimum}
Function
Sets the minimum voltage or current value of the
active selected templet.
Description
When voltage is the object to edit, it is the
minimum voltage to be set. When current is the
object to edit, it is the minimum current to be set.
When templet type is Pulse, the command is to set
high level value.