EXTERNAL CONTROL
179
The LoadOn IN setting determines whether the
load is turned on when the external switch is
closed (low) or open (high).
High
Low
High
Low
LoadOn In = High
Load
On
Off
Load on
Load off
LoadOn In = Low
> Configure [F5] > Next Menu [F4]
> External [F3] and set the LoadOn IN setting.
Set to Low if you want the load to be turned on
when the switch is closed.
Set to High if you want the load to turn on
when the switch is open.
When external control is used to turn the load off, the
load key cannot be used to turn the load on. However
the reverse is not true. If the load has been turned on
by external control, the load key can be used to turn
the load off.