OPERATION
75
Von Delay
Description
Von Delay is the amount of time the unit will wait
before turning the load on after the Von Voltage
threshold has been latched. This will prevent
overshoot current from affecting the Von Voltage
threshold.
Operation
1. Press
Main
> Configure[F5] > Other[F2] and
set the Von Delay time.
Range:
Von Delay: OFF, 2.0-60ms(CC, CV, CP
mode)
OFF, 5.0-60ms (2.0-60ms)(CR
mode)
Note
CR mode can have the delay time set separately from
the other modes (called
Von Delay CR
when in CR
mode).
Timer Functions
Count Time
Description
When Count Time is set to on, it will count the
elapsed time from when the load was turned on to
when it was turned off.
This function is applicable to manual and
automatic shutdown (such as from protection
functions such as UVP etc.)
The elapsed time will be shown in the display
Measurement area.
Operation
1. Press
Main
> Configure[F5] > Other[F2] and
turn the Count Time on or off.
Range:
ON, OFF