PEL-3000(H) User Manual
244
Reverse voltage protection(RVP)
By diode. Turns off the load when an alarm occurs.
Rating overcurrent protection (ROCP)
An ROCP message will be produced when the input current range
is greater than 110% of the rated operating current range (I
range).
Rating overpower protection (ROPP)
An ROPP message will be produced when the input power range
is greater than 110% of the rated operating power range.
Front panel input rating overcurrent protection (F.ROCP)
An F.ROCP message will be produced when the front panel input
current range is greater than 77A (Typical).
Sequence
Normal Sequence
Operation mode
CC, CR, CV or CP
Maximum number of steps
1000
Step Execution Time
0.05ms 999 h 59 min
Time resolution
0.05 ms (0.05 ms 1 min)/100 ms (1 min 1 h)/1 s
(1 h 10 h)/10 s (10 h 100 h)/1 min (100 h 999
h 59 min)
Fast Sequence
Operation mode
CC or CR
Maximum number of steps
1000
Step Execution Time
25 μs 600 ms
Time resolution
1us(25us -60ms) /10us(60.01ms -600ms)
Other
Elapsed Time Delay
Measures the time from load on to load off. On/Off selectable.
Measures from 1 s up to 999 h 59 min 59 s
Auto Load Off Timer
Automatically turns off the load after a specified time elapses.
Can be set in the range of 1 s to 999 h 59 min 59 s or off
Communication Function
GPIB
IEEE std. 488.1-1978 (partial support)
SH1, AH1, T6, L4, SR1, DC1, DT1.
Supports the SCPI and IEEE std. 488.2-1992 command set
Sets panel functions except the power switch and reads measured
values