PEL-3000(H) User Manual
180
Load On/Off Status
Pin 13 (Load On Status) of the J1 connector is used
to monitor the load status (on or off).
The Load On Status pin
is a photo-coupled open-
collector output.
Photocoupler input: 30V max, 8mA, max.
External Control of the Range
The range for the present operating mode can be
externally controlled when the current range is set
to high range.
The range is changed using pins 8, 9 (Range Cont 1
&2) and 12 (A Com) of the J1 connector.
When externally controlling the range, the pin
input combination determines which range is
chosen.
> Configure [F5] > Next Menu
[F4] > External [F3] and set the Control setting
to V, R or Riv to enable external control.
2. When externally controlling the range, the pin
input combination determines which range is
chosen.