OPERATION
71
Display
Mode
CV
I Range
H 8.75A
V Range
L 80V
Response
Slow
Configure
0.000
V
0.0000
A
0.00
W
CV B Value 80.000
CV A Value 80.000
LOADRS232
A Value
Fine
01/Oct/2018
V
V
Response
setting
CC, CR and CP Mode Response Speed
Description
By default, the normal current response” speed is
set to 1/1. The response speed can be reduced to
1/2, 1/5, 1/10.
Reducing the current response speed can affect
other settings such as the slew rate and soft
start settings.
Operation
1. Make sure the load is off.
2. Press
Main
> Configure[F5] > Other[F2] and
set the Response parameter.
Range:
1
1
,
2
1
,
5
1
,
10
1