PEL-3000(H) User Manual
194
Turning the Load On
Operating the PEL-3000(H) Series in parallel mode
is the same as for single units.
When using the units in parallel, the load line
inductance could be increased or the stability of the
units could be reduced. It may be necessary to reduce
the response speed setting to increase stability.
1. Turn the slave and master units on.
2. Set the operation mode and settings on the
master unit.
The master’s settings will be used by the slave
units.
3. Turn the load on from the Master unit.
All measurements will be displayed and
updated on the Master unit only.
Disable Parallel Mode
To disable parallel mode, each unit must be set as a
“Master”.
1. Turn the power off on all the units and remove
the GTL-255 frame link cables.
2. Turn the power back.
3. On each unit, press
> Configure [F5] >
Next Menu [F4] > Parallel[F1].
4. Set the unit to Master with the Operation
setting.
5. Turn the Parallel and Booster settings to Off.