OPERATION
87
OVP
Description
If the OVP is tripped, the PEL-3000(H) will turn off
the load.
The OVP levels can be set from 0V to 10% higher
than the rating voltage.
Operation
1. Press
Main
> Configure[F5] > Protection[F1]
and set the OVP Level.
Range:
OVP Level: OFF, 0-Rating voltage + 10%
Note
To turn OVP off, set the OVP voltage greater than the
current rating voltage + 10%.
Alarm
The OVP indicator and a message will only
appear on the screen when the input voltage is
below the UVP level. The Enter key must be
pressed to clear the alarm message.
To clear the OVP indicator, remove the cause of
the over voltage - i.e., reduce the input voltage.
Display
8.653
V
50.006
mA
0.4327
W
CV B Value 80.000
CV A Value 80.000
LOADRS232
A Value
Fine
01/Oct/2017
V
V
OVP Alarm
Please Press Enter
To Clear Alarm
OVP
OVP indicator
Alarm message
when OVP is
set to Load Off
Mode
CV
I Range
L87.5mA
V Range
L 80V
Response
Fast
Configure