EXTERNAL CONTROL
187
Voltage Monitor Output (PEL-3021H/PEL-3041H/PEL-3111H)
Description
The voltage output from the VMON OUT terminal
and from the VMON pin on the J3 connector is
used to represent the current input level.
The V Range used to represent the full scale
current range from the VMON OUT terminal and
from the VMON pin on the J3 connector depends
on the current range settings.
V MON voltage
Input Voltage
8V or 10V
0V
0V
Rated Voltage
Monitor Connector
Voltage Range
Monitor Output
Range
V MON OUT (BNC)
H, L
0 - 8V
V MON (J3)
H, L
0 - 10V
V MON OUT
BNC Connector
The V MON OUT BNC connector outputs a
voltage of 0 - 8V for the High and Low voltage
ranges. The common potential is connected to the
chassis ground potential.
J3 Connector
The voltage across pins 2 and 3 (or 4) outputs a
voltage of 0 -10V for the High and Low voltage
ranges. The common potential is connected to A
COM (negative load terminal).