PMX command set (API)
PMX A4354-2.2 HBM: public 481
Set analog Output (only PX878)
Analog outputs
Sets the voltage of all selected analog outputs of one or
more PX878 multi IO cards to a given level (-10 V … +10
V). The command activates a test mode and “cuts” the
connection of the analog output from its former‐connec
ted source signal üpath. To deactivate the test mode use
the OSP command.
Syntax: SAO p1 (x)
Parameters: p1: voltage
Example: SAO 1.25(x)
Note: This command causes a lot of CPU load. Setting 10 val
ues per second of a single analog output raises CPU load
by about 15 %.
Important
This command is implemented in PMX firmware 2.00 and
higher.
SAO