Diagnosis and maintenance
520 A4354-2.2 HBM: public PMX
3. User
Copy the "pmxpasswordreset" to the root directory of a
USB stick. Connect this stick to the USB socket of the
PMX system while PMX is in normal operation.
The administrator password will be removed immediately.
The file will be deleted from the USB stick.
You may want to keep a copy of the Reset file in order to
be able to reset the password again in the future. This is
possible with a specific PMX device as long as the host
name is not changed.
24.4 Resetting the PMX to factory settings
All device settings can be reset in the menu Settings >
System > Device > Device administration > Restore
factory settings.
This function is not accessible in user level 1 (Worker).
Loading the factory settings deletes the following
settings:
S All channel and amplifier settings (measurement
channels and calculated channels, e.g. min./max.
values)
S All device settings (e.g. parameter sets)
The following are not deleted:
S The network settings
S The passwords for the various user levels (Worker,
Service, Administrator)