Electromagnetic Immunity Guidance
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The P3700 Affinity® 4 Birthing Bed is intended for use in the electromagnetic environment specified below. The
customer or the user should make sure that it is used in such an environment.
Immunity Test
EN and IEC 60601-1-2
Test Level
EMC Compliance Level
Electromagnetic Environment—
Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
± 6 kV Contact
± 8 kV Air
(see Note 1)
± 8 kV Contact
± 15 kV Air
(see Note 2)
± 6 kV Contact
± 8 kV Air
(see Note 1)
± 8 kV Contact
± 15 kV Air
(see Note 2)
Floors should be wood, concrete,
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Electrical Fast
Transient/Burst
IEC 61000-4-4
± 2 kV for Power Supply
Lines (see Note 1)
5 kHz Repetition
Frequency
± 2 kV for Power Supply
Lines (see Note 2)
100 kHz Repetition
Frequency
± 2 kV for Power Supply
Lines (see Note 1)
5 kHz Repetition
Frequency
± 2 kV for Power Supply
Lines (see Note 2)
100 kHz Repetition
Frequency
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Mains power quality should be that
of a typical commercial or hospital
environment.
Power Frequency
Magnetic Fields
IEC 61000-4-8
3 A/m (see Note 1)
30 A/m (see Note 2)
3 A/m (see Note 1)
30 A/m (see Note 2)
Mains power quality should be that
of a typical commercial or hospital
environment. When the inputs are
greater than 5.5 Vrms, it may dam-
age the speaker system.
Voltage Dips,
Short Interrupts, &
Variations On
Power Supply
Lines
IEC 61000-4-11
(see Notes 1 and
Note 3)
< 5% U
T
(>95% dip in U
T
for 0.5
cycle)
< 40% U
T
(60% dip in U
T
for 5 cycles)
< 70% U
T
(30% dip in U
T
for 25
cycles)
< 5% U
T
(95% dip in U
T
for
5 seconds)
< 5% U
T
(>95% dip in U
T
for 0.5
cycle)
< 40% U
T
(60% dip in U
T
for 5 cycles)
< 70% U
T
(30% dip in U
T
for 25
cycles)
< 5% U
T
(95% dip in U
T
for
5 seconds)
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
P3700 Affinity® 4 Birthing Bed
requires continued operation
during power mains interruptions,
it is recommended that the Affin-
ity® 4 Birthing Bed be powered
from an uninterruptible power
supply or battery.
Voltage Dips
IEC 61000-4-11
(see Notes 2 and
Note 3)
0% U
T
for 0.5 cycle at 0°,
45°, 90°, 135°, 180, 225°,
270°, and 315° (for single-
phase mains equipment);
0% U
T
for 1 cycle
70% U
T
for 25/50 Hz, and
30/60 Hz cycles
0% U
T
for 0.5 cycle at 0°,
45°, 90°, 135°, 180, 225°,
270°, and 315° (for single-
phase mains equipment);
0% U
T
for 1 cycle
70% U
T
for 25/50 Hz, and
30/60 Hz cycles
Voltage
Interruptions
IEC 61000-4-11
(see Notes 2 and
3)
0% U
T
for 250/50 Hz and
300/60 Hz cycles
0% U
T
for 250/50 Hz and
300/60 Hz cycles