3.2 Measuring Insulation Resistance
1
2
3
4
5
6
7
8
9
10
11
付録
索引
63
When the display reflects the following
state, insulation resistance measurement
cannot be started.
The setting value is blinking to indicate
that the instrument being set up
The mark is blinking
•While [TC] is lit, the actual measurement
temperature is shown as [- - -]
An error massage is displayed
If the indication is unstable, the average of
the measurement is shown.
Pressing the average key toggles [AVE] on/
off.
While [AVE] is on, display update interval is
four seconds, normally.
But in the following case, the interval is one
second even if [AVE] is on.
During 15 seconds after the measure-
ment started
During 5 to 10 seconds after the mea-
surement range changed
The state not to be started the measure-
ment
Average function