3.2 Measuring Insulation Resistance
62
10. Read the indication.
If the indication is unstable, press the key.
The average of the measurements is shown.
Resistance indication is switched to leakage current
indication by pressing the key.
See 3.2.5 "Switching to Leakage Current Indication" (page
68).
When the timer has been set, remaining time is dis-
played.
See 4.1 "Using Timer" (page 79).
Do not allow test leads to contact each other
or place objects on test leads, to avoid mea-
surement errors and malfunctions.
Be sure to clean test leads after use. If test
leads are soiled, they may deteriorate.
Insulation resistance is unstable. The
indication may not stabilize with some
objects.
Due to factors such as capacitance of
objects under test, resistance values may
start low, then rise gradually and settle out.
During measurement, if the resistance of
the object suddenly drops or if the test
lead tips are short-circuited, the tester
stops voltage generation as a safety
measure. (This applies to a test voltage of
1.1 kV or more.)