4.3 Temperature Correction (TC)
88
4.3.1 Performing Temperature Correction
Procedure
1. Measure temperature and insulation
resistance. The measurements are held
upon completion.
(Either may be measured first.)
3.2 "Measuring Insulation Resistance" (page 56)
See 3.4 "Measuring Temperature" (page 76).
The temperature may also be entered with
keys.
See 6.3 "Entering Temperature and Humidity
Measured with External Thermometer and
Hygrometer" (page 129).
In the step voltage test mode ([STEP] is on),
temperature correction is unavailable. Exit
the step voltage test mode.
See 4.4.3 "Exiting Step Voltage Test Mode"
(page 99).
2. Press the key.
[TABLE No.] blinks.
3. Choose a table No. from 0 to 9 using the
key.