7.1 General Specifications
1
2
3
4
5
6
7
8
9
10
11
付録
索引
139
Specifications 7
7.1 General Specifications
Operating
temperature
and humidity
0 to 40°C, less than 90%RH (non-con-
densation)
(Battery pack charge: 10 to 40°C, less
than 80%RH)
Storage
temperature
and humidity
-10 to 50°C, less than 90%RH (non-con-
densation)
Battery pack: -20 to 30°C, less than
80%RH (non-condensation)
Period of
guaranteed
accuracy
1 year
Operating
environment
Indoors, Pollution degree 2,
Up to 2000 m (6562ft) ASL
Measuring
method
DC voltage application (insulation resis-
tance) and mean-value rectification (voltage)
A-D conver-
sion
Double integral
Display
LCD with backlight displaying up to a
count of 999
Overflow
indication
>, OF
Underflow
indication
<, -OF
Display update
rate
Insulation resistance/leakage current:
Once/sec. (0.25 times/sec. if averaging
function used)
Output voltage monitor: Twice/sec.
Voltage : Fourth/sec.
Temperature : Once/sec.
Bar graph : Twice/sec.