4.3 Temperature Correction (TC)
1
2
3
4
5
6
7
8
9
10
11
付録
索引
87
• The tester converts measured resistance
to the resistance at a reference tempera-
ture and displays the result.
• There are 10 correction methods (correc-
tion tables) available depending on the
object under test and its characteristics.
Choose the appropriate temperature cor-
rection table.
• The reference temperature may be set to
an arbitrary level. The selectable refer-
ence temperature ranges vary depending
on the correction table used.
• The convertible measurement tempera-
ture ranges also vary depending on the
correction table used.
See Appendix 4 "Temperature Correction Table"
(page 163).
4.3 Temperature Correction (TC)
What is it used for?
Used to acquire insulation resis-
tance at a temperature differing from
the actual temperature at which
measurement is performed.