4.2 Displaying PI and DAR
86
When the resistance indication blinks, the
displayed reading may be incorrect. (Insu-
lation resistance changed rapidly before
end of set specified time, affecting mea-
surement range due to internal circuit fail-
ure to respond)
When the resistance reading blinks, regard
the PI or DAR value as a reference. Per-
form measurement again.
The table below shows special indications
for PI and DAR.
Blinking resistance indication on PI or
DAR display screen
PI, DAR Conditions
- - -
One or more resistance val-
ues could not be acquired.
•([- - -] appears in the resis-
tance field.)
One or more resistance val-
ues exceeded measure-
ment range.
•([
OF] appears in the resis-
tance field.)
The 1st measurement was
0.00 M.
>999
PI or DAR is larger than 999.
<0.01
PI or DAR is smaller than
0.01.