3.3 Hold Functions
37
1
2
3
4
5
6
7
8
9
10
11
Additional Functions
* The threshold value of each function is shown below.
The refresh hold function locks the display if the display
value exceeds the threshold value and the internal mea-
surement value stabilizes to a value within the set range
of threshold values.
The display is updated and locked each time the internal
measurement value exceeds the range of threshold val-
ues for the locked display value.
If the display value does not exceed the threshold value
that was set, the display value is not locked in. If you
have trouble getting the display value to lock, try
changing the threshold value.
Function Threshold value
V 0.05 V
mV 0.5 mV
μA5 μA
mA 0.5 mA
A 0.05 A
Ω OL
±Threshold
value
Display updated
and locked
A single buzzer sounds
Connected to
test subject
Disconnect
Display locked
Threshold
value
Internal
measured
value
Display
value locked
Display
value locked
Display locked
A single buzzer
sounds
Connected to
test subject
Disconnect
Display locked