52
4.2 Toggling the Percentage Display (4 - 20mA/0 - 20mA)
When measuring with the DCmA function, press the
DUAL key three times to display the percentage display in
the sub display.
This procedure sets the percentage display method.
4.2 Toggling the Percentage Display
(4 - 20mA/0 - 20mA)
4-20 mA Converts the current over a range of 4 mA to 20
mA into a percentage from 0 to 100%.
0-20 mA Converts the current over a range of 0 mA to 20
mA into a percentage from 0 to 100%.
34
1. In order to display the Power On
Option setting screen, hold dow
n
the
SHIFT
key while turning th
e
function switch.
2. Press the
HOLD
key 5 times to di
s-
play the Percentage Display se
t-
ting screen.
3. Press the
RANGE
key or the
DUA
L
key to select the percentage di
s-
play setting.
4. Confirm the selection by pressin
g
the
Hz
key.
5. Hold down the
SHIFT
key to ent
er
the measurement screen.
Otherwise, turn the function switc
h
to OFF to finish setup.
4-20 mA
(default)
/ 20 mA
1
SHIFT +
Power On
2
(5 times)
5
Select through
power on option
Percentage
conversion value =
Measured value[mA] - 4[mA]
16[mA]
X 100[%]
Measured value[mA]
20[mA]
X 100[%]
Percentage
conversion value =