Measurement
Measurement 2
Observe the following precautions to avoid electric
shock.
Always verify the appropriate setting of the func-
tion selector before connecting the test leads.
Disconnect the test leads from the measurement
object before switching the function selector.
The maximum input voltage is 1000 VDC, 1000
Vrms (sin) or 10
7
VHz. Attempting to measure volt-
age in excess of the maximum input could destroy
the instrument and result in personal injury or
death.
The maximum input current is as follows;
A terminal: Continuous up to 10 A AC/DC, no more
than 30 seconds up to 20 A AC/DC.
mA.mA terminal: 510 mA AC/DC
Never exceed this limit, as doing so could result in
destruction of the instrument and personal injury
or death.
The maximum rated voltage between input termi-
nals and the ground is as follows;
CAT III: DC1000 V, AC1000 Vrms (sin) or 10
7
VHz
CAT IV: DC600 V, AC600 Vrms (sin) or 10
7
VHz
Attempting to measure voltages exceeding this
level with respect to ground could damage the
instrument and result in personal injury.
To avoid electrical shock, be careful to avoid
shorting live lines with the test leads.
For safety, test lead connections must always be
made at the secondary side of a circuit breaker.