4.7 Refresh Hold Setting
This procedure sets the threshold value for the refresh
hold function. The threshold value locks the display to
showing the amount of change after the display value has
stabilized, using the set amount of change as a criteria.
Setting this to off disables the refresh hold function and
enables the trigger hold function.
4.7 Refresh Hold Setting
34
1. In order to display the Power On
Option setting screen, hold dow
n
the
SHIFT
key while turning th
e
function switch.
2. Press the
REL
key 5 times to di
s-
play the Refresh Hold settin
g
screen.
3. Set the Refresh Hold function.
RANGE
key: Increases the value
DUAL
key: Decreases the value
4. Confirm the selection by pressin
g
the
Hz
key.
5. Hold down the
SHIFT
key to ent
er
the measurement screen.
Otherwise, turn the function switc
h
to OFF to finish setup.
OFF (default)
/ 100 counts to 1000
counts
1
SHIFT +
Power On
2
(5 times)
5