4.8 Auto Power Save Function
59
1
2
3
4
5
6
7
8
9
10
11
Power On Options
This procedure sets the time until the auto power save
function is activated. The auto power save function is
activated after the set time elapses since the last time the
rotary switch or a key was operated. The auto power
save function turns off the display screen and minimizes
the instrument's internal power consumption.
4.8 Auto Power Save Function
• In order to resume operation from the auto power
save function, either turn the rotary switch to OFF
and turn the power back on again or else press any
key. In the case of pulse output, press any of the
HOLD, REL, RANGE or DUAL keys or set the function
switch to a position other than pulse output and then
back to its original position. The 3802-50 reverts to
the power ON state.
• The auto power save function is automatically dis-
abled when either the peak hold function or the
recording function is used.
• To disable the auto power save function, hold down
the key while turning the function switch to turn on
the instrument.
34
1. In order to display the Power O
Option setting screen, hold dow
the
SHIFT
key while turning th
function switch.
2. Press the
REL
key 4 times to di
play the Auto Power Save settin
screen.
3. Set the Auto Power Save functio
RANGE
key: Increases the value
DUAL
key: Decreases the value
SHIFT
key: Selects the digit
4. Confirm the selection by pressin
the
Hz
key.
5. Hold down the
SHIFT
key to ent
the measurement screen.
Otherwise, turn the function switc
to OFF to finish setup.
15 min (default)
/ 1 to 99 min
/ OFF
1
SHIFT +
Power On
2
(4 times)
5