70
5.2 Accuracy
Ω (Resistance measurement)/ Contact Check
Diode
C (Capacitance Measurement)
Range Accuracy
Measured
Current
Open Terminal
Voltage
Overload
Protection
510.00 Ω ±0.08%rdg.±10 dgt.
*1
Approx.
1.00 mA
4.8 V
1000 V DC/
1000 Vrms
(sin) or
10
7
VHz,
for 1 minute
Overload cur-
rent:
0.3 A or less
5.1000 kΩ ± 0.08%rdg. ± 5 dgt.
*1
Approx.
0.38 mA
51.000 kΩ
± 0.08%rdg. ± 5 dgt.
Approx.
38 μA
510.00 kΩ
Approx.
3.8 μA
5.1000 MΩ ± 0.2%rdg. ± 5 dgt.
Approx.
345 nA
51.000 MΩ ± 1%rdg. ± 5 dgt.
*2
Approx.
200 nA
510.00 nS ± 1%rdg. ± 10 dgt.
*3
Sampling rate: 7.5 times/ s
Continuity threshold value: Buzzer sounds less than 1000 counts for each range
*1: Accuracy is specified after the inputs are shorted and the relative value (REL)
display function has been initiated
*2: Specified for humidity up to 60%RH
*3: 50 nS or less
Range Accuracy
Measured
Current
Open Terminal
Voltage
Overload
Protection
2.1000 V ± 0.1%rdg. ± 5 dgt.
Approx.
1.00 mA
4.8 V
1000 V DC/
1000 Vrms
(sin) or 10
7
VHz,
for 1 minute
Sampling rate: 7.5 times/ s
Continuity threshold value: Buzzer sounds at less than 0.0500 V
Range Accuracy
Sampling Rate
(Maximum)
Overload Protection
9.999 nF ± 2.5%rdg. ± 8 dgt.
4 times/ s
1000 V DC/ 1000 Vrms (sin)
or 10
7
VHz, for 1 minute
Overload current:
0.3 A or less
99.99 nF
± 1.5%rdg. ± 5 dgt.
999.9 nF
9.999 μF
99.99 μF
999.9 μF once/ s
9.999 mF 0.1 time/ s
99.99 mF ± 3.5%rdg. ± 10 dgt. 0.01 time/ s