SecurView DX Workstation V12.0 Quality Control Manual
Chapter 3: QC Activities for the Radiologic Technologist
MAN-09466 Revision 001 Page 11
After completing the QC procedure, do the following for Selenia Dimensions systems only:
1. At the Acquisition Workstation, select Admin > Quality Control > Technologist tab
> Diagnostic Review Workstation Quality Control.
2. Select Mark Completed to label the status of this procedure as finished.
3. To mark the QC procedure as completed, select Yes.
3.2 Corrective Action
If any display does not meet the performance criteria, adjust room lighting conditions
and repeat the action that failed. If the action fails after the second attempt, correct the
problem before performing mammographic image interpretation using the instructions
below.
If any performance criterion is not met, use the Auto-calibration procedure to recalibrate
the display. Then rerun the Hologic Compliance Test and verify that the display
performs within the recommended limits.
A qualified Service Engineer may have to resolve any issues if display operating levels
are not within recommended ranges.
If the details of a test contain no results, the test shall be run manually after making sure
the displays are correctly attached and turned on. A qualified Service Engineer may have
to resolve any issues if the details of a test contain no results after the second attempt.
While you resolve display calibration issues, SecurView systems can still receive images.
Use an alternative diagnostic device (for example, a second diagnostic workstation or
DICOM printer) for mammography image interpretation until all issues are resolved and
the displays perform within recommended ranges.