EasyManua.ls Logo

Hologic SecurView DX - 6.2 Scheduling the MediCal QAWeb Hologic Compliance Test

Hologic SecurView DX
54 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SecurView DX Workstation V12.0 Quality Control Manual
Chapter 6: Additional MediCal QAWeb Procedures
MAN-09466 Revision 001 Page 39
6.2 Scheduling the MediCal QAWeb Hologic Compliance Test
By default, the Hologic Compliance Test automatically runs weekly (except on systems
using external photometer sensor displays). Use this procedure to change the time/day of
the scheduled Hologic Compliance Test.
1. Log out of the SecurView DX application.
2. To start MediCal QAWeb, press the Windows
key, then select Menu > Barco >
MediCal QAWeb Agent.
3. The MediCal QAWeb Agent window opens. Select the Status icon.
4. The Workstation > Status window opens. Select Info next to ‘Hologic Compliance
Test’.
5. The Hologic Compliance Test window opens. Select Modify next to ‘Frequency’.
6. The MediCal QAWeb login window opens. By default, the User is Advanced’. Type
advanced (case sensitive) in the Passwordfield, then select OK.
7. The Modify Hologic Compliance Test Frequency window opens. Select Custom from the
Frequency: drop-down menu, then select Properties.
8. The Modify Custom Frequency of Hologic Compliance Test window opens. In the
‘Timing’ section, select the test start time, then select the day(s) you want the test to
run. Select OK.
9. At the Modify Hologic Compliance Test Frequency window, select OK.
10. Close all open windows.

Table of Contents

Other manuals for Hologic SecurView DX

Related product manuals