SecurView DX Workstation V12.0 Quality Control Manual
Chapter 4: Testing the Displays
MAN-09466 Revision 001 Page 27
e. Select View Detailed Result. A window opens with the detailed results.
Figure 16: Hologic Compliance Test – View Detailed Result Window
f. Record the value for Action Result in
Appendix B: SecurView Weekly QC Form for
Radiologic Technologists on page 43.
Important
If the details of a test contain no results, follow the instructions in section Corrective
Action on page 11.
Note
You may Print the results or Save as PDF for a local copy from this window. Select
Close once you are finished.
g. Select Close to return to the Hologic Compliance Test window.
h. Close all open MediCal QAWeb windows.