SecurView DX Workstation V12.0 Quality Control Manual
Chapter 4: Testing the Displays
Page 24 MAN-09466 Revision 001
c. Verify that the results meet the performance criteria.
d. Record the results in the
Appendix A: SecurView Annual QC Form for Medical
Physicists on page 41.
e. Select Close twice to return to the Workstation > Status window.
f. Close all open MediCal QAWeb windows.
To test the displays:
1. Log out of the SecurView DX application.
2. To start MediCal QAWeb, press the Windows
key, then select Menu > Barco >
MediCal QAWeb Agent.
3. The MediCal QAWeb Agent window opens. Select the Status icon.
Figure 13: MediCal QAWeb Agent Window