SecurView DX Workstation V12.0 Quality Control Manual
Chapter 4: Testing the Displays
MAN-09466 Revision 001 Page 29
c. Verify that the results meet the performance criteria.
d. Record the results in the
Appendix A: SecurView Annual QC Form for Medical
Physicists on page 41.
e. Select Close twice to return to the Workstation > Status window.
f. Close all open MediCal QAWeb windows.