EasyManua.ls Logo

Honeywell GasAlertMicro 5 - Page 94

Honeywell GasAlertMicro 5
105 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
82
GasAlertMicro 5/PID/IR
User Manual
Table 21. Troubleshooting
Problem Possible Cause Solution
Alarms Troubleshooting
Detector does not enter
alarm mode.
Alarm setpoint(s) are set
incorrectly.
Reset alarm setpoints. Refer to Calibration and Setting Alarm Setpoints
.
Alarm setpoint(s) are set to
zero.
Reset alarm setpoints. Refer to Calibration and Setting Alarm Setpoints
.
Detector requires calibration. Calibrate the sensors. Refer to Calibration and Setting Alarm Setpoints
.
Detector intermittently
enters alarm without any
apparent reason.
Ambient gas levels are near
alarm setpoint or the sensor is
exposed to a puff of the target
gas.
Detector is operating normally. Use caution in suspect areas. Check MAX
gas exposure reading.
Alarms setpoints are set
incorrectly.
Reset alarm setpoints. Refer to Calibration and Setting Alarm Setpoints
.
Missing or faulty sensor. Refer to Replacing a Sensor or Sensor Filter
.
Detector enters alarm with
LEL reading displaying as
OL (over limit).
Pump flow rate is set too high. Refer to the last solution in Pump Operation troubleshooting on the
following page.
Automatic Deactivation Troubleshooting
Detector automatically
deactivates.
Automatic shutdown activated
because of weak batteries.
Refer to Charging the Battery
.
Calibration is overdue and the
Due-lock (calibration user
lock-out) option is enabled.
Enter the password to enter calibration. Calibrate the sensor(s). Refer to
Calibration Procedure
.

Table of Contents

Related product manuals