Event Codes
MAN0984_Iss 4_01/19 Touchpoint Plus
Pt. No. 3011M5001 149 Technical Handbook
Event
Type
Code
Long Description
Triggered
Display in Event
History
Details
Fault
13
Auxiliary battery fault
Main Board
Battery Fault
[Enclosure]
Enclosure
0: master enclosure,
1: extension enclosure
Fault
14
IO board not matched or present
Main Board
IO B/D Not Matched
Slot: xx-x, Prev: xxx,
Curr: xxx
Prev, Curr B/D type
0:Main, 1:UI, 2:mA
Input, 3 : mV input,
4:Relay, 5:mA output,
7:Modbus,
7:Extension power
Slot
Refer to the note 1
Fault
15
Cell is producing a negative reading
(depending on gas under-range limit)
Main Board
Negative Reading
CH: xx, Conc: xxxx
[unit]
Fault
16
Fault self-cleared
Main Board
Fault Cleared
CH: xx, Code: xx
Code
Code Number (1~17)
Fault
17
Reset all faults by a user
Main Board
Reset all faults
Inhibit
1
Inhibited input channel by LUI
Main Board
Inhibited input
channel
CH: xx
Inhibit
2
Cleared inhibit mode for input channel by
LUI
Main Board
Cleared inhibit mode
CH: xx
Inhibit
3
Inhibited all input channels remotely
Main Board
Inhibited all channels
remotely
Inhibit
4
Cleared inhibit mode for all input channel
remotely
Main Board
Cleared inhibit mode
remotely
CH: xx
Inhibit
5
Detected inhibit input from detector
IO board
Detector In Inhibit
CH: xx
Warning
1
Battery low (not critical)
Main Board
Battery is low
[Enclosure], xxx %
Enclosure
0: master enclosure,
1: extension enclosure
Warning
2
Battery is not being charged
Main Board
Battery not charged
[Enclosure]
Enclosure
0: master and
extension enclosure
Warning
3
RESERVED
Warning
4
Over-range warning
(depending on gas over-range limit)
Main Board
Over range
CH: xx, Conc: xxxx
[unit]
Warning
5
Calibration is due soon
Main Board
Calibration is
overdue
CH: xx, Due: xxxx
Due
Calibration interval -
days passed after
successful calibration
Warning
6
Time/date not set
UI Board
Date/Time not set
[Date/Time]