5
Figure 5-2. The Status Registers
The OPERation branch of the status registers indicates things that can happen
during normal operation. The QUEStionable branch indicates error conditions.
:STATus:OPERation:CONDition?
Syntax
:STATus:OPERation:CONDition?
Description
This query reads the contents of the OPERation:CONDition
register. Only two bits of the condition register are used:
Bit 8, which is 1 when the output power that you set exceeds
what the laser is capable of producing.
Bit 9, which is 1 while the instrument is performing its
power-up initialization.
Example
OUTPUT 724;":STAT:OPER:COND?"
ENTER 724;A$
5-30 Remote Commands