5
[:SOURCE]:MODOUT
Note
This command does not conrm to the SCPI standard.
Syntax
[
:SOURCE
]
:MODOUT FRQ
j
FRQRDY
j
0
j
1
Description
This command sets the modulation output. The two possible
values are
FRQ
or
0
, where the modulation signal is output all the time,
and
FRQRDY
or
1
, where the modulation signal is combined with
the laser-ready signal, so that the output is kept low when
there is no optical signal being output (for example
, while the
laser is settling after a change of wavelength).
[:SOURCE]:MODOUT?
Note
This command does not conrm to the SCPI standard.
Syntax
[
:SOURCE
]
:MODOUT?
Description
This queries the mode of the modulation output. The two
possible values are
0
, where the modulation signal is output all the time, and
1
, where the modulation signal is combined with the
laser-ready signal, so that the output is kept low when there
is no optical signal being output (for example, while the laser
is settling after a change of wavelength).
Remote Commands 5-21