the instrument cabinet. The type of Mains plug shipped with each
instrument depends on the country of destination. Refer to Figure 3-2
for the part numbers of the power cables and mains plugs available.
Power Requirements
The signal generator requires a power source of either 50/60/400 Hz
at 100/120
V
or 50/60 Hz at 200/240
V.
The voltage ranges for these
nominal voltage values
are
shown in Bble 3-1. Power consumption is
150
VA
maximum.
"hble
3-1.
Voltage Ranges
for
Nominal Voltage Values
I
Available
ac
Voltage
1
Voltage Range
I
90
to
110
V
rms
108 to 132
V
rms
198 to 242
V
rms
216 to 250
V
rms
Caution
This instrument has autoranging line voltage input; be sure the supply
voltage is within the specified range.
Replacing the Fuse
If
a fuse failure is suspected, replace the 250
V,
3A, type
F
fuse (HP
part number 2110-0780)
as
follows:
1. Unplug the power cord from the line module.
2. Use a flat-bladed screw driver to pry and unseat the fuse housing
3. Remove the cartridge and inspect the fuse positioned toward the
from the line module.
front of the instrument.
3-2
Installation