1.2. DURING USE
Read the following recommendations and instructions carefully:
CAUTION
Failure to comply with the warnings and/or instructions may result in
damage to the equipment and/or its components or personal injury to the
operator
Do not touch any unused test terminal
Do not disconnect the cords during measurement
Do not open the battery compartment when the instrument inputs are
connected
The symbol indicates that the internal batteries are fully charged. When
the charge level drops to the minimum level, the symbol appears on the
display. In this case, stop the tests and replace or recharge the batteries as
described in § 9.2
The instrument will not charge the batteries if the temperature is
outside the permitted range of them (0°C ÷ 40 °C / 32°F ÷ 104°F)
The instrument retains data even without batteries
1.3. AFTER USE
When measurements are complete, switch off the instrument by pressing and holding the
ON/OFF key for some seconds. If the instrument is not to be used for a long period,
remove the batteries and follow the instructions in § 3.3.
1.4. DEFINITION OF MEASUREMENT CATEGORY (OVERVOLTAGE)
Standard IEC/EN61010-1: Safety requirements for electrical equipment for measurement,
control and laboratory use, Part 1: General requirements”, defines what measurement
category, commonly called overvoltage category, is. In § 6.7.4: Measuring circuits, it states
that:
Circuits are divided into the following measurement categories:
Measurement Category IV is for measurements performed at the source of the low-
voltage installation.
Examples are electricity meters and measurements on primary overcurrent protection
devices and ripple control units.
Measurement Category III is for measurements performed in the building installation
Examples are measurements on distribution boards, circuit breakers, wiring, including
cables, bus-bars, junction boxes, switches, socket-outlets in the fixed installation, and
equipment for industrial use and some other equipment, for example, stationary motors
with permanent connection to fixed installation.
Measurement Category II is for measurements performed on circuits directly
connected to the low voltage installation.
Examples are measurements on household appliances, portable tools and similar
equipment.
Measurement Category I is for measurements performed on circuits not directly
connected to MAINS.
Examples are measurements on circuits not derived from MAINS, and specially
protected (internal) MAINS-derived circuits. In the latter case, transient stresses are
variable; for that reason (OMITTED) the user must know the transient withstand
capability of the equipment