6.5.3.1. Meaning of measurement results
In general, the result of a test on a Voc and Isc measurement is determined by the
following relationships. Please note the following parameters:
= open circuit voltage rating @STC
= short-circuit current rating @STC
= Positive tolerance value on Voc = x
= Negative tolerance value on Voc = x
= Positive tolerance value on Isc = x
= Negative tolerance value on Isc = x
= Maximum instrument error declared on Voc (see §10.1 )
= Maximum instrument error declared on Isc (see § 10.1)
= Error on Voc measurement @ STC
= Error on Isc measurement @ STC
The following conditions on parameters on the measurement result are managed by the
instrument:
(1) → Positive test result also considering instrument error in the measurement
(2) → Positive test result, except in case of instrument error in the measurement
(3) → Negative test result except in case of instrument error in the measurement
(4) → Negative test result also considering instrument error in the measurement