3-14
i-CAT FLX Installation Manual
032-0330-EN Rev K
Scenario 3 - New DEXIS - i-CAT FLX Installation
1
Install i-CAT FLX (Chapters 1 and 2)
Scanner and controller installed and networked with IP address.
SmartScan STUDIO is pre-installed at the factory.
2
Install DEXIS Core / DEXIS Eleven
Install and Configure DEXIS Core:
Using the DEXIS Core media, install DEXIS Core on the server that functions as the long-term
data storage device. Refer to the DEXIS Eleven Software Manual for instructions.
Install DEXIS Eleven client software:
Using the DEXIS Eleven media, install DEXIS client software on all clinical workstations that are
using DEXIS. DEXIS Eleven client software can also be installed on the server that functions as
the long-term data storage device on a shared server-client. Refer to the DEXIS Eleven
Software Manual for instructions.
3
Install SmartScan STUDIO Server Components (page 3-17) on long-term storage server.
Integration Services (required by all sites)
PACS Storage Service (sites using PACS)
Data Utility (optional)
Patient Data Utility (optional)
Legacy Practice Management Interface (sites using PM system)
DEXIS i-CAT FLX Server Plugin (required for all DEXIS sites)
NOTE: For a Shared Server - Client installation scenario, you must also load the DEXIS i-CAT
FLX Client on the server (located on Client Components media). For a Separate Server - Client
installation scenario, load the DEXIS i-CAT FLX Client component on the server only if the site
plans to use the server as a client.
4
Install SmartScan STUDIO Client Components (page 3-21) on workstations in front office,
operatories, and consultation rooms.
SmartScan STUDIO Manager (required on one workstation for accessing QA test images)
PACS QR (sites using a PACS for image storage)
PACS MWL (sites using modality worklists for scheduling patients)
Legacy Practice Management Viewer Launcher (sites using PM system)
DEXIS i-CAT FLX Client (required for all DEXIS sites)
5
Install InVivoDental Software (page 3-24) as required.
If not already loaded, load on the workstation where the site will be viewing QA Tests.
Load on workstations that will be using it to view patient images, as purchased by the site.
NOTE: Vision is not supported on the i-CAT FLX, only TxSTUDIO and Dolphin viewers are
supported.