Clarity Ultrasonic Surgical Aspirator System Operator’s Manual
11-10
Disassembling the System
Detaching the Contamination Guard
NOTE
The contamination guard should not be removed from the console
unless it is time to replace it (every six months or when the color
changes).
1. Detach the handpiece cable from the console by gently pulling on the
handpiece cable connector in the handpiece port on the front of the
console.
2. Discard the contamination guard in compliance with hospital policy for
contaminated waste.
Powering Down the System
Press the Power button on the front of the console. The touchscreen will
go dark. Do not unplug power cord until system has powered down.