CUSA
®
Clarity Ultrasonic Surgical Aspirator System Operator’s Manual
11-11
Cleaning the System
NOTE
Do not wipe the touch screen with a sponge because it may scratch the
surface.
NOTE
Never apply cleaner directly to the touch screen.
WARNING
Do not allow uids to enter the console.
WARNING
Electric Shock Hazard - Always unplug the CUSA Clarity System
before cleaning it.
CAUTION
Product damage may result if you do not follow the instructions
described in the cleaning section.
Cleaning the System
Cleaning the Touchscreen
Make sure to the console is unplugged before cleaning it. Clean the touch
screen with a lint-free wipe and a cleaner, such as window/glass cleaner,
eyeglass cleaner, or a quaternary ammonia disinfectant.
Cleaning the Console and Cart
1. Unplug the power cord from the wall receptacle and the console.
2. Clean the surface with a lint-free wipe soaked with one of the
following:
• 70% IPA
• Super Sani-Cloth
®
• Sani-Cloth
®
Bleach
3. Make sure the surfaces remain visibly wet for a minimum of two (2)
minutes, then air dry.