CUSA
®
Clarity Ultrasonic Surgical Aspirator System Operator’s Manual
7-5
Assembling the Handpiece in the Sterile Field
Attaching the Nosecone and the Flue
1. Slide the larger end of the nosecone over the tip and slide it down over
the neck of the handpiece making sure the white dots are aligned.
2. Rotate the nosecone clockwise until it locks into place and the white
dot on the nosecone aligns with the while dot on the body of the
handpiece.
Make sure the white dots are aligned on the nosecone and handpiece.
3. Slide the ue onto the nosecone base along the tip.
Remove stylet from the ue and put it aside in the sterile eld
if needed to clear the tip during surgery.