CUSA
®
Clarity Ultrasonic Surgical Aspirator System Operator’s Manual
9-5
Testing the Handpiece
2. If the test fails:
a. Remove the tip and then re-attach with the torque wrench and
base, and re-run the test. If the error continues to persist, attach
a new tip with the torque wrench and base, and re-run the test.
b. If the error continues to persist after replacing the tip, replace the
entire handpiece.
c. Re-run the test.
3. If the test passes, press the Main Screen button.
After testing the system, a Details button displays. Press this button to
view the following detailed test results.
Detailed Test Results
• Tip Frequency Optimization - Finds the best frequency for the tip
• Amplitude Test - Measures how close the handpiece is to maximum
amplitude and provides a score from 0 to 100%. The score must be
greater than or equal to 50% to pass the test
Press the Main Screen button to begin surgical use.
For a description of the settings, see “Section 4 Touchscreen Display
and Functions” on page 4-1.
System Ready for Use