CUSA
®
Clarity Ultrasonic Surgical Aspirator System Operator’s Manual
10-6
Changing a Tip
Assembling the New Tip on the Handpiece
1. Open new sterile tip pack.
2. Thread a tip onto the handpiece connecting body by turning the tip
clockwise until it is nger tight.
3. Place the torque base on a at surface or in the designated spot in the
sterilization tray.
4. Place the handpiece on the torque base as shown and make sure that
the 2 at metal sides t snugly in the metal slot at the end of the base.
5. Hold the handpiece in the base and place the provided torque wrench
(color side towards handpiece) over the tip.