Clarity Ultrasonic Surgical Aspirator System Operator’s Manual
A-5
WARNING
Explosion Hazard - Do not use the CUSA Clarity System in the
presence of ammable anesthetics or other volatile solvents.
WARNING
To avoid risk of electric shock, this equipment must only be connected to
a supply mains with protective earth.
Environment
NOTE
Other devices in the operating room may generate electromagnetic
interference. Use caution in locating equipment within the room to
reduce the electromagnetic interference.
Maximum Low Frequency Leakage
Touch current <500µA
Patient leakage current <50µA
Under maximum loading conditions, the CUSA Clarity console is suitable
for ultrasonics activation times of 10 minutes on, 5 minutes off.
The system requires a minimum of one hour exposure at its operating
temperature range before you use it.
See the EMC Information section, page A-6 for further information on
environmental conditions.
Electromagnetic Interference
The CUSA Clarity system console minimizes electromagnetic interference
to other equipment used in the operating room. The system complies with
the requirements of IEC 60601-1-2: 2014 Fourth Edition.
Duty Cycle
Environment
Operating
Temperature range 10°C (50°F) to 35°C (95°F)
Humidity range
30 to 85%, relative humidity,
non-condensing
Atmospheric pressure range
70 kPa to 106 kPa (10.15 to
15.38 psi)
Moisture
Console Ingress Protection Rating IPX1
Storage and Shipping
Temperature range -20°C (-4°F) to 60°C (140°F)
Humidity range
15 to 85% relative humidity,
non-condensing