CUSA
®
Clarity Ultrasonic Surgical Aspirator System Operator’s Manual
13-2
Alarm Descriptions
NOTE
The external USB drive should be a FAT, FAT32, exFAT or NTFS
formatted drive. Other le formats are not supported.
NOTE
When the extraction is complete, a message to remove the USB drive
displays.
Extracting Log Files
Alarm Descriptions
1. From the Main Screen, press the Settings button .
2. Press System Information.
3. Attach a USB drive to the USB connector on the rear of the console.
4. Press the Extract Log button.
5. Remove the USB drive.
Condition Cause Recommendation
Low Aspiration
The console has
measured the
aspiration level to be
too low to be
functioning properly.
• Check tubing
connections and
routing
• Replace tubing set
if problem persists
Handpiece Failure
A fault has occurred
reading or writing to
the handpiece
memory chip.
• Disconnect and
reconnect
handpiece
• Replace handpiece
if problem persists
Footswitch Failure
A footswitch wiring or
connection fault has
been detected.
• Disconnect and
reconnect
footswitch
• Replace footswitch
if problem persists
Console Overheating
A high temperature
has been measured
within the console.
The console will
automatically shut
off.
Contact Integra
Service