CUSA
®
Clarity Ultrasonic Surgical Aspirator System Operator’s Manual
11-13
Cleaning the System
NOTE
Make sure to keep the connector dry.
NOTE
Do not clean the sterilizer tray with abrasives. Product damage will
result.
Cleaning the Footswitch
1. Disconnect the footswitch from the console.
2. Wipe it clean.
3. If the footswitch is contaminated with blood or uid:
a. Immerse the footswitch in warm water containing detergent or
disinfectant.
b. Rinse the footswitch with clean water.
4. Allow the footswitch to drain after rinsing.
Sterilization Tray Cleaning
Use a neutral detergent to clean the tray.