CUSA
®
Clarity Ultrasonic Surgical Aspirator System Operator’s Manual
7-4
Assembling the Handpiece in the Sterile Field
3. Hold the handpiece in the base and place the torque wrench
(color side towards handpiece) over the tip.
4. Rotate the torque wrench clockwise until it clicks twice.
NOTE
Keep the torque wrench and base in the sterile eld in case you need to
change a tip during the procedure and for disassembly after surgery.
CAUTION
To avoid product damage, always use the torque base to hold the
handpiece while using the torque wrench to tighten or loosen the tip.
Avoid over torqueing the tip.