Clarity Ultrasonic Surgical Aspirator System Operator’s Manual
8-5
Console Setup (Non-Sterile)
WARNING
If the aspiration tubing and contamination guard are not connected to
the correct canister ports, the system may not work.
3. Close the cartridge door.
4. Connect the contamination guard to the vacuum port of the canister.
Make sure the contamination guard is in place and check expiration
date.
NOTE
Contamination guard must be replaced every six months or when the
color changes.
NOTE
Since there are several brands of canisters, the canister may appear
different than in the picture above.