Clarity Ultrasonic Surgical Aspirator System Operator’s Manual
10-8
Changing a Tip
9. Slide the ue onto the nosecone base along the tip.
10. Make sure that the ue is aligned with the pre-aspiration holes at the
distal end of the tip.
NOTE
If the pre-aspiration holes are completely covered, most of the irrigation
uid may be aspirated back through the handpiece, impairing cavitation,
thus reducing tissue ablation efcacy.
Remove stylet from the ue and put it aside in the sterile eld if
needed to clear the tip during surgery.