CUSA
®
Clarity Ultrasonic Surgical Aspirator System Operator’s Manual
10-10
Changing a Tip
12. Push the irrigation tubing into the clip at the base of the handpiece.
NOTE
If needed, the clip can be rotated to minimize strain on the user’s
wrist.
NOTE
Handpiece cable is not intended to be inserted into the clip at the base
of the handpiece.
13. Prime the irrigation and test the handpiece before starting to use the
handpiece again by pressing “start prime” and “start test”.